Home > Fully Automated Double Beam Atomic Absorption Spectrophotometer  
Fully Automated Double Beam Atomic Absorption Spectrophotometer  
Fully Automated Double Beam Atomic Absorption Spectrophotometer (AAS4141)
 
• Fully automatic selection of wave length, lamp current, PMT gain adjustment, data handling and storage of results through a PC.
• Double beam Ebert type monochromator with 1800 lines/mm holographic grating, 0 to 2nmSBW, 185 to 930nm wave length range with resolution <0.2nm.
• The full wave length range of monochromator is motorized for rapid scanning.
• Automatic motorized turret for operation with 4 lamps.
• Visible display of the signal energy in the form of bar signal and numerical display.
• Automatic safety inter locks provided for mains voltage, air, flame sensor, liquid trap, flash back and fuel supply.
• Automatic calibration using up to maximum 10 standards.
• Sensitivity >0.7 absorbance for 5ppm aqueous solution of Copper with Air-Acetylene flame.
• Fitted with special deigned Burner assembly and Gas control Units.
 
Application:
 
AAS4141 systems are used for quantitative analysis of trace elements.